Electrocardiographic markers for septal stimulation of the right ventricular outflow tract

Authors

Keywords:

cardiac pacing artificial, ventricular septum, electrocardiogram.

Abstract

Introduction: Electrocardiographic patterns of true septal stimulation of the right ventricular outflow tract are not clearly identified.

Objective: To identify electrocardiographic markers of septal stimulation of the right ventricular outflow tract.

Methods: Observational study in 30 patients with permanent ventricular electrode implantation in the right ventricular outflow tract. For the 12-lead electrocardiogram, stimulation was programmed in VVI mode at 100 beats per minute. The position of the electrode was validated with two-dimensional transthoracic echocardiography.

Results: 21/30 (80%) were true septal implants. The mean QRS duration was shorter in the septal group (126.2 ± 9.7 ms vs. 144.4 ± 15.1 ms; p = 0.003) and the QRS notching in DII, DIII, aVL and aVF was related to non-septal location; p = 0.005, p = 0.019, p = 0.049 and p = 0.005, respectively. The net QRS in DII (14.3 ± 6.0 mm vs. 9.3 ± 3.0 mm; p = 0.045), DIII (12.6 ± 5.8 mm vs. 8.0 ± 4.7 mm; p = 0.036) and aVF (14.0 ± 6.4 mm vs. 8.8 ± 4.3 mm; p = 0.025) was higher in the septal group. The duration of the stimulated QRS ≤ 135 ms showed a sensitivity of 81.0 % and a specificity of 66.7% and the absence of notching in inferior leads, a sensitivity of 95.2% and a specificity of 44.4% to identify septal stimulation.

Conclusions: Electrocardiographic markers of septal stimulation of the right ventricular outflow tract are heterogeneous; validity indicators, QRS duration ≤ 135 ms and the absence of notching in inferior leads, may be adequate indicators of right ventricular outflow septal stimulation.

Downloads

Download data is not yet available.

References

Laksono S, Setianto B, Iqbal M, Prawara AS. Understanding Pacemaker-Induced Cardiomyopathy Incidence and Predictors in Patients with Right Ventricular Pacing: A Systematic Review. Int J Angiol. 2021;31(1):10-15. DOI: https://doi.org/10.1055/s-0041-1735206 .

Mizner J, Jurak P, Linkova H, Smisek R, Curila K. Ventricular Dyssynchrony and Pacing-induced Cardiomyopathy in Patients with Pacemakers, the Utility of Ultra-high-frequency ECG and Other Dyssynchrony Assessment Tools. Arrhythm Electrophysiol Rev. 2022;11:e17. DOI: https://doi.org/10.15420/aer.2022.01

Dawood M, Elsharkawy E, Abdel-Hay MA, Nawar M. Predictors of pacing induced left ventricular dysfunction and cardiomyopathy assessed by three-dimensional echocardiography and speckle tracking strain. Egypt Heart J. 2021;73(1):10. DOI: https://doi.org/10.1186/s43044-021-00136-x

Cruz M, Cort Z, Gutiérrez A, Mengana A, Ramos MY. Estimulación septal del ventrículo derecho como alternativa a la estimulación apical. ¿Cómo ubicar el electrodo? Rev Cuban Cardiol Cir Cardiovasv. 2020 [acceso 03/10/2021];26(3):1561-2937. Disponible en: http://www.revcardiologia.sld.cu/index.php/revcardiologia/article/view/914

Bansal R, Parakh N, Gupta A, Juneja R, Naik N, Yadav R, et al. Incidence and predictors of pacemaker-induced cardiomyopathy with comparison between apical and non-apical right ventricular pacing sites. J Interv Card Electrophysiol. 2019;56(1):63-70. DOI: https://doi.org/10.1007/s10840-019-00602-2

Squara F, Scarlatti D, Riccini P, Garret G, Moceri P, Ferrari E. Classical fluoroscopy criteria poorly predict right ventricular lead septal positioning by comparison with echocardiography. J Interv Card Electrophysiol. 2018;52(2):209-15. DOI: https://doi.org/10.1007/s10840-018-0355-x

Kaye GC, Rowe MK, Gould PA. Using the Surface ECG to Identify Right Ventricular Pacing Lead Position: A Cautionary Tale. Pacing Clin Electrophysiol. 2017;40(9):1039-41. DOI: https://doi.org10.1111/pace.13065

Burri H, Park CI, Zimmermann M, Gentil-Baron P, Stettler C, Sunthorn H, et al. Utility of the surface electrocardiogram for confirming right ventricular septal pacing: validation using electroanatomical mapping. Europace. 2011;13(1):82-6. DOI: https://doi.org/10.1093/europace/euq332

Rowe MK, Moore P, Pratap J, Coucher J, Gould PA, Kaye GC. Surface ECG and Fluoroscopy are Not Predictive of Right Ventricular Septal Lead Position Compared to Cardiac CT. Pacing Clin Electrophysiol. 2017;40(5):537-44. DOI: https://doi.org/10.1111/pace.13066

Kusumoto FM, Schoenfeld MH, Barrett C, Edgerton JR, Ellenbogen KA, Gold MR, et al. 2018 ACC/AHA/HRS Guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2019;74(7):e51-e156. DOI: https://doi.org/10.1016/j.jacc.2018.10.044 .

Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, et al. Corrigendum to: 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC): With the special contribution of the European Heart Rhythm Association (EHRA). Europace 2022;24(4):699. DOI: https://doi.org/10.1093/europace/euac023

Mǎrgulescu AD, Suran BM, Rimbaş RC, Dulgheru RE, Silişte C, Vinereanu D. Accuracy of fluoroscopic and electrocardiographic criteria for pacemaker lead implantation by comparison with three-dimensional echocardiography. J Am Soc Echocardiogr. 2012;25(7):796-803. DOI: https://doi.org/10.1016/j.echo.2012.04.010 .

Das A, Kahali D. Ventricular septal pacing: Optimum method to position the lead. Indian Heart J 2018;70(5):713-20. DOI: https://doi.org/10.1016/j.ihj.2018.01.023

Shenthar J, Rai MK, Chakali SS, Pillai V, Delhaas T. Computed tomography validated right ventricular mid-septal lead implantation using right ventricular angiography. J Arrhythm. 2021;37(5):1131-8. DOI: https://doi.org/10.1002/joa3.12591

Squara F, Scarlatti D, Riccini P, Garret G, Moceri P, Ferrari E. Individualized Left Anterior Oblique Projection: A Highly Reliable Patient-Tailored Fluoroscopy Criterion for Right Ventricular Lead Positioning. Circ Arrhythm Electrophysiol. 2018;11(7):e006107. DOI: https://doi.org/10.1161/CIRCEP.117.006107

Wei H, Tang J, Chen D, Zhang Q, Liang Y, Liu L, et al. Electrocardiographic predictors of validated right ventricular outflow tract septal pacing for correct localization of transthoracic echocardiography. Clin Cardiol. 2018;41(3):354-9. DOI: https://doi.org/10.1002/clc.22873

Israel CW, Tribunyan S, Yen Ho S, Cabrera JA. Anatomy for right ventricular lead implantation. Herzschrittmacherther Elektrophysiol. 2022;33(3):319-26. DOI: https://doi.org/10.1007/s00399-022-00872-w

Enriquez A, Baranchuk A, Briceno D, Saenz L, García F. How to use the 12-lead ECG to predict the site of origin of idiopathic ventricular arrhythmias. Heart Rhythm. 2019;16(10):1538-44. DOI: https://doi.org/10.1016/j.hrthm.2019.04.002

Hahn RT, Waxman AB, Denti P, Delhaas T. Anatomic relationship of the complex tricuspid valve, right ventricle, and pulmonary vasculature: A review. JAMA Cardiol. 2019;4(5):478-87. DOI: https://doi.org/10.1001/jamacardio.2019.0535

Mala A, Osmancik P, Herman D, Curila K, Stros P, Vesela J, et al. Can QRS morphology be used to differentiate between true septal vs. apparently septal lead placement? An analysis of ECG of real mid-septal, apparent mid-septal, and apical pacing. Eur Heart J Suppl. 2020;22(F):14-22. DOI: https://doi.org/10.1093/eurheartj/suaa094

Dixit S, Gerstenfeld EP, Callans DJ, Marchlinski FE. Electrocardiographic patterns of superior right ventricular outflow tract tachycardias: distinguishing septal and free-wall sites of origin. J Cardiovasc Electrophysiol 2003;14(1):1-7. DOI: https://doi.org/10.1046/j.1540-8167.2003.02404.x

McGavigan AD, Roberts-Thomson KC, Hillock RJ, Stevenson IH, Mond HG. Right ventricular outflow tract pacing: radiographic and electrocardiographic correlates of lead position. Pacing Clin Electrophysiol. 2006;29(10):1063-8. DOI: https://doi.org/10.1111/j.1540-8159.2006.00499.x

Hillock RJ, Stevenson IH, Mond HG. The right ventricular outflow tract: a comparative study of septal, anterior wall, and free wall pacing. Pacing Clin Electrophysiol. 2007;30(8):942-7. DOI: https://doi.org/10.1111/j.1540-8159.2007.00790.x

Shimeno K, Yoshiyama T, Abe Y, Akamatsu K, Kagawa S, Matsumoto R, et al. The usefulness of right ventriculography to aid anchoring a pacing lead to the right ventricular septum. Europace. 2018;20:1154‐60. DOI: https://doi.org/10.1093/europace/eux165

Pang BJ, Joshi SB, Lui EH, Tacey MA, Ling LH, Alison J, et al. Validation of conventional fluoroscopic and ECG criteria for right ventricular pacemaker lead position using cardiac computed tomography. Pacing Clin Electrophysiol. 2014;37(4):495-504. DOI: https://doi.org/10.1111/pace.12301

Gupta A, Parakh N, Bansal R, Verma SK, Roy A, Sharma G, et al. Correlation of pacing site in right ventricle with paced QRS complex duration. Indian Pacing Electrophysiol J. 2018;18(6):210-16. DOI: https://doi.org/10.1016/j.ipej.2018.08.001

Matsuo M, Shimeno K, Yoshiyama T, Matsumura Y, Matsumoto R, Abe Y, et al. Utility of the combination of simple electrocardiographic parameters for identifying mid-septal pacing. J Cardiovasc Electrophysiol. 2019;30(11):2433-40. DOI: https://doi.org/10.1111/jce.14174

Pastore G, Zanon F, Baracca E, Rigatelli G, Corbucci G, Mazza A, et al. How can we identify the optimal pacing site in the right ventricular septum? A simplified method applicable during the standard implanting procedure. Am J Cardiovasc Dis. 2013;3(4):264-72.

Kamakura S, Shimizu W, Matsuo K, Taguchi A, Suyama K, Kurita T, et al. Localization of optimal ablation site of idiopathic ventricular tachycardia from right and left ventricular outflow tract by body surface ECG. Circulation. 1998;98(15):1525-33. DOI: https://doi.org/10.1161/01.cir.98.15.1525

Osmancik P, Stros P, Herman D, Curila K, Petr R. The insufficiency of left anterior oblique and the usefulness of right anterior oblique projection for correct localization of a computed tomography-verified right ventricular lead into the midseptum. Circ Arrhythm Electrophysiol. 2013;6(4):719-25. DOI: https://doi.org/10.1161/CIRCEP.113.000232

Chen D, Wei H, Tang J, Liu L, Wu S, Lin C, et al. A randomized comparison of fluoroscopic techniques for implanting pacemaker lead on the right ventricular outflow tract septum. Int J Cardiovasc Imaging. 2016;32(5):721-8. DOI: https://doi.org/10.1007/s10554-016-0840-1

Balt JC, van Hemel NM, Wellens HJ, de Voogt WG. Radiological and electrocardiographic characterization of right ventricular outflow tract pacing. Europace. 2010;12(12):1739-44. DOI: https://doi.org/10.1093/europace/euq341

Shima T, Ohnishi Y, Inoue T, Yoshida A, Shimizu H, Itagaki T, et al. The relation between the pacing sites in the right ventricular outflow tract and QRS morphology in the 12-lead ECG. Jpn Circ J. 1998;62(6):399-404. DOI: https://doi.org/10.1253/jcj.62.399

Andrikopoulos G, Tzeis S, Asbach S, Semmler V, Lennerz C, Solzbach U, et al. SPICE Study Investigators. A stepwise electrocardiographic algorithm for differentiation of mid-septal vs. apical right ventricular lead positioning: the SPICE ECG substudy. Europace. 2015;17(6):915-20. DOI: https://doi.org/10.1093/europace/euu344

Published

2022-12-21

How to Cite

1.
Cruz Cardntey M, Cort Z, Mengana Betancourt A, Blanco Pérez Y, Soto García Z, Gutierrez Lopez A. Electrocardiographic markers for septal stimulation of the right ventricular outflow tract. Rev. cuba. cardiol. cir. cardiovasc. [Internet]. 2022 Dec. 21 [cited 2025 Feb. 12];28(4):e1405. Available from: https://revcardiologia.sld.cu/index.php/revcardiologia/article/view/1405

Issue

Section

Original Articles

Most read articles by the same author(s)

1 2 > >>